> The example on the left shows a printout with an ISO/GLP
header, which is only printed in the manual data transfer
mode.
The last digit of the weight must be printed inversely
because the verification scale interval (e) of the weighing
instrument used in legal metrology is not equal to its
actual scale interval (d).
> After data transfer, the printer transmits a signal to the
weighing instrument so that the instrument will be tared.
§ Leave the tablet on the weighing pan
§ Add the next tablet to the one already on the weighing pan
> The weight is automatically read into the statistics memory
when the weighing instrument has stabilized
> The next weight is printed out (example)
> The weighing instrument is tared
When the weighing instrument displays a zero readout:
§ Add the next tablet to the weighing pan
§ Press the key
§ As described above, add the next tablet each time after the
weighing instrument has been tared
When you have placed the last tablet on the weighing pan
and the last weight is displayed and printed:
§ Generate a statistical evaluation by pressing the key
> The statistical evaluation will be printed (in this case,
only
those values preset in the “Statistics Applications”
menu:
number of weighing operations, mean value, sum of the
individual weights, minimum and maximum weights)
§ Press the c key on the weighing instrument
> The ISO/GLP footer will be printed
123456789012345678901234
SARTORIUS
Model BP6100
S/N 040240772
Id
Date : 22-May-95
Start: 10:25:14
Ser. :
001: + 9.81 g
123456789012345678901234
002: + 10.13 g
123456789012345678901234
n 12
X 9.98 g
x 120.34 g
min 9.60 g
max 10.32 g
End : 10:31:56
Name :